ISO/TS 10974:2012 is applicable to implantable parts of active implantable medical devices (AIMDs) intended to be used in patients who might undergo a magnetic resonance scan in 1,5T, cylindrical bore, whole body MR scanners for imaging the hydrogen nucleus.

The tests that are specified in ISO/TS 10974:2012 are type tests intended to be carried out on samples of a device to characterize interactions with the magnetic and electromagnetic fields associated with an MR scanner.

ISO/TS 10974:2012 contains test methods that are applicable to a broad class of AIMDs for the purpose of evaluating device operation against several hazards.